产品描述
The product CSB-E11789h is a sandwich ELISA kit developed to measure levels of human tryptophanyl-tRNA synthetase (WARS) in serum, plasma, cell lysates, or tissue homogenates. The enzyme-substrate chromogenic reaction is also used to amplify the signal and quantify the levels of the analyte through the intensity of the colored product. The color intensity positively correlates with the amount of WARS bound in the initial step.
WARS is an essential enzyme because it catalyzes the ligation of tryptophan (Trp) to its cognate tRNAtrp during translation via aminoacylation. It is localized not only to the nucleus but also to the extracellular space, playing an important role in innate immunity, angiogenesis, and IFN-γ signaling. WARS expression varies significantly in different tissues and pathological states, implying that it plays unique roles in physiological homeostasis and immune defense. It is also involved in the pathogenesis of multiple diseases, including sepsis, cancer, and autoimmune and brain diseases.
别名
Gamma 2 ELISA Kit; GAMMA-2 ELISA Kit; Gamma2 ELISA Kit; hWRS ELISA Kit; IFI 53 ELISA Kit; IFI53 ELISA Kit; IFP 53 ELISA Kit; IFP53 ELISA Kit; Interferon induced protein 53 ELISA Kit; Interferon-induced protein 53 ELISA Kit; SYWC_HUMAN ELISA Kit; T2-TrpRS ELISA Kit; TrpRS ELISA Kit; Tryptophan tRNA ligase; cytoplasmic ELISA Kit; Tryptophan tRNA ligase 1 cytoplasmic ELISA Kit; Tryptophan tRNA ligase ELISA Kit; Tryptophan--tRNA ligase ELISA Kit; Tryptophanyl tRNA synthetase ELISA Kit; Tryptophanyl tRNA synthetase cytoplasmic ELISA Kit; WARS ELISA Kit; WARS protein ELISA Kit; WRS ELISA Kit
缩写
WARS
Uniprot No
P23381
灵敏度
0.078 ng/mL
反应时间
1-5h
样本体积
50-100ul
检测波长
450 nm
研究领域
Epigenetics and Nuclear Signaling
测定原理
quantitative
精密度
| Intra-assay Precision (Precision within an assay): CV%<8% |
|
|
|
| Three samples of known concentration were tested twenty times on one plate to assess. |
|
| Inter-assay Precision (Precision between assays): CV%<10% |
|
|
|
| Three samples of known concentration were tested in twenty assays to assess. |
|
|
| |
|
|
|
|
|
|
线性度
| To assess the linearity of the assay, samples were spiked with high concentrations of human WARS in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. |
| |
Sample |
Serum(n=4) |
|
| 1:1 |
Average % |
91 |
|
| Range % |
88-95 |
|
| 1:2 |
Average % |
87 |
|
| Range % |
81-93 |
|
| 1:4 |
Average % |
97 |
|
| Range % |
93-101 |
|
| 1:8 |
Average % |
100 |
|
| Range % |
94-106 |
|
回收率
| The recovery of human WARS spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. |
| Sample Type |
Average % Recovery |
Range |
|
| Serum (n=5) |
89 |
84-95 |
|
| EDTA plasma (n=4) |
93 |
88-99 |
|
| |
|
|
|
|
|
|
| |
|
|
|
|
|
|
标准曲线
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. |
| ng/ml |
OD1 |
OD2 |
Average |
Corrected |
|
| 20 |
1.972 |
2.050 |
2.011 |
1.920 |
|
| 10 |
1.446 |
1.556 |
1.501 |
1.410 |
|
| 5 |
1.027 |
1.101 |
1.064 |
0.973 |
|
| 2.5 |
0.638 |
0.606 |
0.622 |
0.531 |
|
| 1.25 |
0.381 |
0.387 |
0.384 |
0.293 |
|
| 0.625 |
0.293 |
0.275 |
0.284 |
0.193 |
|
| 0.312 |
0.195 |
0.201 |
0.198 |
0.107 |
|
| 0 |
0.093 |
0.089 |
0.091 |
|
|
|
货期
3-5 working days