收藏| Intra-assay Precision (Precision within an assay): CV%<8% | |||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | |||||||
| Inter-assay Precision (Precision between assays): CV%<10% | |||||||
| Three samples of known concentration were tested in twenty assays to assess. | |||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of rabbit S-100B in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | |||||||
| Sample | Serum(n=4) | ||||||
| 1:1 | Average % | 88 | |||||
| Range % | 84-92 | ||||||
| 1:2 | Average % | 94 | |||||
| Range % | 90-99 | ||||||
| 1:4 | Average % | 103 | |||||
| Range % | 99-110 | ||||||
| 1:8 | Average % | 93 | |||||
| Range % | 89-98 | ||||||
| The recovery of rabbit S-100B spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | |||||||
| Sample Type | Average % Recovery | Range | |||||
| Serum (n=5) | 95 | 90-100 | |||||
| EDTA plasma (n=4) | 97 | 92-104 | |||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||