This Guinea pig TNF ELISA Kit was designed for the quantitative measurement of Guinea pig TNF protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 3.12 pg/mL-200 pg/mL and the sensitivity is 0.78 pg/mL .
别名
TNF ELISA Kit; TNFA ELISA Kit; TNFSF2 ELISA Kit; Tumor necrosis factor ELISA Kit; Cachectin ELISA Kit; TNF-alpha ELISA Kit; Tumor necrosis factor ligand superfamily member 2 ELISA Kit; TNF-a) [Cleaved into: Tumor necrosis factor ELISA Kit; membrane form ELISA Kit; N-terminal fragment ELISA Kit; NTF); Intracellular domain 1 ELISA Kit; ICD1); Intracellular domain 2 ELISA Kit; ICD2); C-domain 1; C-domain 2; Tumor necrosis factor ELISA Kit; soluble form] ELISA Kit
缩写
TNF
Uniprot No
P51435
灵敏度
0.78 pg/mL
反应时间
1-5h
样本体积
50-100ul
检测波长
450 nm
研究领域
Immunology
测定原理
quantitative
精密度
| Intra-assay Precision (Precision within an assay): CV%<8% | | | |
| Three samples of known concentration were tested twenty times on one plate to assess. | |
| Inter-assay Precision (Precision between assays): CV%<10% | | | |
| Three samples of known concentration were tested in twenty assays to assess. | | |
| | | | | | | |
线性度
| To assess the linearity of the assay, samples were spiked with high concentrations of guinea pig TNF-α in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. |
| | Sample | Serum(n=4) | |
| 1:1 | Average % | 96 | |
| Range % | 90-100 | |
| 1:2 | Average % | 94 | |
| Range % | 89-98 | |
| 1:4 | Average % | 92 | |
| Range % | 87-96 | |
| 1:8 | Average % | 93 | |
| Range % | 89-100 | |
回收率
| The recovery of guinea pig TNF-α spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. |
|
| Sample Type | Average % Recovery | Range | |
| Serum (n=5) | 92 | 88-97 | |
| EDTA plasma (n=4) | 97 | 92-103 | |
| | | | | | | |
| | | | | | | |
标准曲线
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. |
|
| pg/ml | OD1 | OD2 | Average | Corrected | | | 200 | 2.350 | 2.376 | 2.363 | 2.200 | | | 100 | 1.884 | 1.902 | 1.893 | 1.730 | | | 50 | 1.368 | 1.379 | 1.374 | 1.211 | | | 25 | 0.961 | 0.973 | 0.967 | 0.804 | | | 12.5 | 0.664 | 0.681 | 0.673 | 0.510 | | | 6.25 | 0.475 | 0.486 | 0.481 | 0.318 | | | 3.12 | 0.305 | 0.309 | 0.307 | 0.144 | | | 0 | 0.162 | 0.164 | 0.163 | | |
|
货期
3-5 working days