收藏| Intra-assay Precision (Precision within an assay): CV%<15% | ||||||
| Three samples of known concentration were tested twenty times on one plate to assess. | ||||||
| Inter-assay Precision (Precision between assays): CV%<15% | ||||||
| Three samples of known concentration were tested in twenty assays to assess. | ||||||
| To assess the linearity of the assay, samples were spiked with high concentrations of mouse INH-A in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. | ||||||
| Sample | Serum(n=4) | |||||
| 1:1 | Average % | 85 | ||||
| Range % | 81-89 | |||||
| 1:2 | Average % | 92 | ||||
| Range % | 87-98 | |||||
| 1:4 | Average % | 105 | ||||
| Range % | 101-109 | |||||
| 1:8 | Average % | 87 | ||||
| Range % | 83-90 | |||||
| The recovery of mouse INH-A spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. | ||||||
| Sample Type | Average % Recovery | Range | ||||
| Serum (n=5) | 84 | 80-87 | ||||
| EDTA plasma (n=4) | 101 | 98-103 | ||||
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. | |||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||