产品描述
This Human GDNF ELISA Kit was designed for the quantitative measurement of Human GDNF protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 78 pg/mL-5000 pg/mL and the sensitivity is 19.5 pg/mL.
别名
Astrocyte derived trophic factor ELISA Kit; Astrocyte derived trophic factor 1 ELISA Kit; Astrocyte-derived trophic factor ELISA Kit; ATF 1 ELISA Kit; ATF 2 ELISA Kit; Atf ELISA Kit; ATF1 ELISA Kit; ATF2 ELISA Kit; gdnf ELISA Kit; GDNF_HUMAN ELISA Kit; Glial cell derived neurotrophic factor ELISA Kit; Glial Cell Line Derived Neurotrophic Factor ELISA Kit; Glial cell line-derived neurotrophic factor ELISA Kit; Glial derived neurotrophic factor ELISA Kit; HFB1 GDNF ELISA Kit; hGDNF ELISA Kit; HSCR3 ELISA Kit
缩写
GDNF
Uniprot No
P39905
灵敏度
19.5 pg/mL
反应时间
1-5h
样本体积
50-100ul
检测波长
450 nm
研究领域
Neuroscience
测定原理
quantitative
精密度
| Intra-assay Precision (Precision within an assay): CV%<8% |
|
|
|
| Three samples of known concentration were tested twenty times on one plate to assess. |
|
| Inter-assay Precision (Precision between assays): CV%<10% |
|
|
|
| Three samples of known concentration were tested in twenty assays to assess. |
|
|
| |
|
|
|
|
|
|
线性度
| To assess the linearity of the assay, samples were spiked with high concentrations of human GDNF in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. |
| |
Sample |
Serum(n=4) |
|
| 1:1 |
Average % |
93 |
|
| Range % |
89-95 |
|
| 1:2 |
Average % |
101 |
|
| Range % |
97-105 |
|
| 1:4 |
Average % |
92 |
|
| Range % |
85-96 |
|
| 1:8 |
Average % |
97 |
|
| Range % |
91-100 |
|
回收率
| The recovery of human GDNF spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. |
| Sample Type |
Average % Recovery |
Range |
|
| Serum (n=5) |
95 |
90-98 |
|
| EDTA plasma (n=4) |
97 |
93-100 |
|
| |
|
|
|
|
|
|
| |
|
|
|
|
|
|
标准曲线
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. |
| pg/ml |
OD1 |
OD2 |
Average |
Corrected |
|
| 5000 |
2.066 |
1.952 |
2.009 |
1.848 |
|
| 2500 |
1.856 |
1.776 |
1.816 |
1.655 |
|
| 1250 |
1.542 |
1.523 |
1.533 |
1.372 |
|
| 625 |
1.122 |
1.137 |
1.130 |
0.969 |
|
| 312.5 |
0.709 |
0.684 |
0.697 |
0.536 |
|
| 156.25 |
0.478 |
0.459 |
0.469 |
0.308 |
|
| 78 |
0.296 |
0.304 |
0.300 |
0.139 |
|
| 0 |
0.163 |
0.159 |
0.161 |
|
|
|
货期