产品描述
This Human HLA-G ELISA Kit was designed for the quantitative measurement of Human HLA-G protein in serum, plasma, tissue homogenates. It is a Sandwich ELISA kit, its detection range is 1.25 ng/mL-80 ng/mL and the sensitivity is 0.31 ng/mL .
别名
B2 microglobulin ELISA Kit; DADB-15K14.8 ELISA Kit; HLA 6.0 ELISA Kit; HLA class I histocompatibility antigen alpha chain G ELISA Kit; HLA class I histocompatibility antigen; alpha chain G ELISA Kit; HLA class I molecule ELISA Kit; HLA G ELISA Kit; HLA G antigen ELISA Kit; HLA G histocompatibility antigen class I G ELISA Kit; HLA G3 ELISA Kit; HLA-G ELISA Kit; HLA-G histocompatibility antigen; class I ELISA Kit; HLA60 ELISA Kit; HLAG ELISA Kit; HLAG_HUMAN ELISA Kit; Major histocompatibility complex class I G ELISA Kit; MHC class I antigen ELISA Kit; MHC class I antigen G ELISA Kit; MHC G ELISA Kit; T-cell A locus ELISA Kit; TCA ELISA Kit
缩写
HLA-G
Uniprot No
P17693
灵敏度
0.31 ng/mL
反应时间
1-5h
样本体积
50-100ul
检测波长
450 nm
研究领域
Immunology
测定原理
quantitative
精密度
| Intra-assay Precision (Precision within an assay): CV%<8% | | | |
| Three samples of known concentration were tested twenty times on one plate to assess. | |
| Inter-assay Precision (Precision between assays): CV%<10% | | | |
| Three samples of known concentration were tested in twenty assays to assess. | | |
| | | | | | | |
线性度
| To assess the linearity of the assay, samples were spiked with high concentrations of human HLA-G in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. |
| | Sample | Serum(n=4) | |
| 1:1 | Average % | 93 | |
| Range % | 86-98 | |
| 1:2 | Average % | 103 | |
| Range % | 97-108 | |
| 1:4 | Average % | 93 | |
| Range % | 85-98 | |
| 1:8 | Average % | 97 | |
| Range % | 91-101 | |
回收率
| The recovery of human HLA-G spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. |
|
| Sample Type | Average % Recovery | Range | |
| Serum (n=5) | 86 | 80-94 | |
| EDTA plasma (n=4) | 96 | 92-104 | |
| | | | | | | |
| | | | | | | |
标准曲线
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. |
|
| ng/ml | OD1 | OD2 | Average | Corrected | | | 80 | 2.961 | 2.960 | 2.961 | 2.828 | | | 40 | 2.252 | 2.205 | 2.229 | 2.096 | | | 20 | 1.266 | 1.286 | 1.276 | 1.143 | | | 10 | 0.731 | 0.755 | 0.743 | 0.610 | | | 5 | 0.368 | 0.400 | 0.384 | 0.251 | | | 2.5 | 0.302 | 0.322 | 0.312 | 0.179 | | | 1.25 | 0.227 | 0.215 | 0.221 | 0.088 | | | 0 | 0.134 | 0.132 | 0.133 | | |
|
货期
3-5 working days