The product CSB-EL001419HU is a sandwich ELISA kit developed to measure levels of human Afamin (AFM) in multiple samples, including serum, plasma, or tissue homogenates. The enzyme-substrate chromogenic reaction is also used to amplify the signal and quantify the levels of the analyte through the intensity of the colored product. The color intensity positively correlates with the amount of AFM bound in the initial step.
AFM is a human plasma vitamin E-binding glycoprotein mainly produced in the liver. It promotes vitamin E transport across the blood-brain barrier. AFM is also a tumor marker for ovarian cancer. AFM has been found strongly associated with all components of metabolic syndrome including visceral obesity and has been a potential early marker of metabolic syndrome. It also plays a role in anti-apoptotic cellular processes associated with oxidative stress and is linked to insulin resistance and other features of metabolic syndrome.
别名
AFAM_HUMAN ELISA Kit; Afamin ELISA Kit; Afm ELISA Kit; ALB 2 ELISA Kit; ALB2 ELISA Kit; ALBA ELISA Kit; ALF ELISA Kit; Alpha albumin ELISA Kit; Alpha-Alb ELISA Kit; Alpha-albumin ELISA Kit
缩写
AFM
Uniprot No
P43652
灵敏度
0.97 ng/mL
反应时间
1-5h
样本体积
50-100ul
检测波长
450 nm
研究领域
Signal Transduction
测定原理
quantitative
精密度
| Intra-assay Precision (Precision within an assay): CV%<8% | | | |
| Three samples of known concentration were tested twenty times on one plate to assess. | |
| Inter-assay Precision (Precision between assays): CV%<10% | | | |
| Three samples of known concentration were tested in twenty assays to assess. | | |
| | | | | | | |
线性度
| To assess the linearity of the assay, samples were spiked with high concentrations of human AFM in various matrices and diluted with the Sample Diluent to produce samples with values within the dynamic range of the assay. |
| | Sample | Serum(n=4) | |
| 1:1000 | Average % | 96 | |
| Range % | 90-102 | |
| 1:2000 | Average % | 89 | |
| Range % | 84-95 | |
| 1:4000 | Average % | 100 | |
| Range % | 95-106 | |
| 1:8000 | Average % | 97 | |
| Range % | 92-104 | |
回收率
| The recovery of human AFM spiked to levels throughout the range of the assay in various matrices was evaluated. Samples were diluted prior to assay as directed in the Sample Preparation section. |
|
| Sample Type | Average % Recovery | Range | |
| Serum (n=5) | 89 | 93-96 | |
| EDTA plasma (n=4) | 102 | 98-107 | |
| | | | | | | |
| | | | | | | |
标准曲线
| These standard curves are provided for demonstration only. A standard curve should be generated for each set of samples assayed. |
|
| ng/ml | OD1 | OD2 | Average | Corrected | | | 250 | 2.024 | 2.098 | 2.061 | 1.929 | | | 125 | 1.715 | 1.761 | 1.738 | 1.606 | | | 62.5 | 1.334 | 1.364 | 1.349 | 1.217 | | | 31.25 | 0.893 | 0.884 | 0.889 | 0.757 | | | 15.6 | 0.567 | 0.546 | 0.557 | 0.425 | | | 7.8 | 0.321 | 0.349 | 0.335 | 0.203 | | | 3.9 | 0.261 | 0.246 | 0.254 | 0.122 | | | 0 | 0.131 | 0.133 | 0.132 | | |
|
货期